Scope calibrator


Posted on Jan 28, 2013

The calibrator operates on exactly 100 kHz providing a reference for calibrating the variable time base oscillator of general purpose scopes. For example, if the scope is set so that one cycle of the signal fills exactly 10 graticule divisions then each division represents 1 MHz, or 1 microsecond. If the scope is adjusted for 10 cycles on 10 graticule divisions.


Scope calibrator
Click here to download the full size of the above Circuit.

(1 cycle per division) then each division represents 100 kHz or 10 microseconds.